overview amino acid analyzer ion chromatography hpcpc applications hpcpc hpcpc accessories various detectors column ovens, degassers off. sample preparation on. sample preparation auto. sample preparation valves column selector online injector & sampler centrion ws autosampler encon p & t concentrator markelov autosampler est cobra autosampler hplc pumps syringe pumps flavourspec breathspec bp-ims gc-ims analytical ims uv-ims µims-odour ims-odour
 
 
 
 
HPCPC SYSTEM         CHROMATOGRAPHY SYSTEM
 
 
High Performance Centrifugal Partition Chromatography
 
SPECIFICATIONS
Model CPC 240 CPC 1400
Rotational Speed
Washing Speed
Separation Speed
0 - 2000 rpm (100 rpm steps)
300 rpm
1,500 rpm
0 - 2000 rpm (100 rpm steps)
300 rpm
1,100 rpm
Flow Rate
Washing Flow
Separation Flow
0.1 - 30 ml per minute
20 ml per minute
5 ml per minute
1 - 200 ml per minute
80 ml per minute
25 ml per minute
Standard Valves 4-way mode switching valve 4-way mode switching valve
6-way sample injection valve
Display LCD display LCD display
Roter Material Poly-phenylenesulfide (PPS)
Stainless Steel
Poly-chlorotrifluoroethylene (DAIFLON)
Stainless Steel
Number of Disk 12 pieces x 1 pack 18 pieces x 1 pack
Rotor Dimension 200Ø xH 80 mm 300Ø xH 150 mm
Rotor Volume 240 ml 1,400 ml
Partition Cells 2136 1296
Cell Length 15 mm 29 mm
Centrifugal Radius 82.5 mm average 120 mm average
Estimated Separation/ Run 24 ml/ run 140 ml/ run
Maximum Pressure

60 kg/cm©÷

60 kg/cm©÷

Safety Device The rotor is automatically stopped when its lip is opened.
In/ Out Line Tubing Size 1/16" PEEK (0.75 ID x 1 mm OD) 1/16" PEEK (0.75 ID x 1 mm OD)

Rotor Net Weight

7 KG 25 KG
Dimension W 330 x D 475 x H 480 mm W 420 x D 610 x H 500 mm
Weight 43 KG 70 KG
 
 
OPTIONAL CPC CONTROLLER SPECIFICATIONS
Executed Files 10 files is exclusively used for washing only
Sequential separation by means of a same file
Most suitable conditions can be determined by using linkage of multiple files
Sequential injections into same stationary phase
Valves Valve 1: Maximum 6 different solvents can be inter-switched
Valve 2: 6-way valve for injection and loading
Valve 3: Ascend and Descend modes can be switched
Display LCD, 40 characters x 4 lines
Function keys: F1, F2, F3 and F4
New Weight 17 KG
Dimension W 380 x D 350 x H 500 mm
 
OPTIONAL SOLVENT DELIVERY PUMP SPECIFICATIONS
Flow Rate 0.1 - 50 ml per minute
Maximum Pressure 7.0 Mpa during connection with CPC
Plunger Diameter 7.0 Ø
Display Pressure, Flow Rate, Upper/ Lower Pressure Limit (Pre-settable)
Net Weight 5 KG
Dimension W 80 x D 310 x H140 mm
The optional items are dedicated controller, semi-preparative pump, injector, UV-Vis detector and fraction collector.
 
 